原子发射双谱线测温系统设计

更新时间:2020-07-24 21:38:36 大小:2M 上传用户:守着阳光1985查看TA发布的资源 标签:测温系统 下载积分:5分 评价赚积分 (如何评价?) 打赏 收藏 评论(0) 举报

资料介绍

为了解决爆炸场反应剧烈、产生的瞬时温度高、作用时间短、环境十分恶劣,导致传统的接触式测温方法不再适用的问题,设计了1套瞬态高温测试系统,该系统基于原子发射光谱双谱线测温法的原理,利用Y型光纤,将爆炸场的光谱辐射信号传入测量系统;选用2条波长间隔小的谱线Cu I 510.5.5nm和Cu I 521.8nm,2条谱线的波长间隔小,大大减小了光谱辐射率、光谱透射率等对光谱测量的影响,提高了测温精度。实验中对热电偶测温法、比色测温法、原子发射光谱双谱线测温法进行了对比,结果表明原子发射光谱双谱线测温法测温上限高,测温精度好,相对误差小于1%。

The traditional contact-type temperature measurement method is no longer applicable to these new situations of the vio-lent reactions of explosion field, the effects of the instantaneous high temperature, the short time and the bad environment nowa-days. A set of transient high temperature measured system was designed to solve the problem by applying the double line of atom-ic emission spectroscopy, which could pass explosion field spectral radiation signals to the measurement system through using the Y type optical fiber. Two spectral lines are Cu I 510. 5 nm and Cu I 521. 8 nm respectively. The effect of the spectral radiance and spectral transmissivity on the spectra...